This portable steam sterilizer has a single drum formed in one piece from thick stainless steel, with no inner seams to hold residue. School, college and small testing labs use it to sterilise glassware, metal instruments and culture media, and to treat used cultures before washing. It is simple to run and economical to use.
| Parameter | Detail |
|---|---|
| Type | Portable autoclave, single-drum design |
| Drum | Formed in one piece from thick stainless steel, no inner joints |
| Operation | Very simple to operate |
| Running | Economical to use |
| Purpose | Steam sterilisation for routine laboratory work |
| Product code | SLE/LA/03 |
| Minimum order | 10 units |
| Dispatch | Usually within 7 to 10 working days |
| Manufacturer | ScienceLab (Jain Scientific Equipments Pvt Ltd), Ambala, India |
Plate pouring practical. Make up nutrient agar in a conical flask with screw cap or in McCartney bottles, with each cap backed off half a turn. Wrap glass Petri dishes in paper, put a strip of autoclave tape on each bundle and run everything through one cycle. Pour the plates once the agar has cooled enough to hold the flask comfortably.
Making used cultures safe. Collect spent plates and tubes in an autoclave bag, leave the neck loosely closed so steam can get in, and stand the bag in an autoclave basket inside the drum. After the cycle and full cooling, the treated waste goes for disposal and reusable glass goes for washing.
Instruments for dissection. Wrap forceps and the blades from a student dissecting set in paper, mark each pack with tape and add them to the next load. Students then compare packs that went through the cycle with packs that did not, which opens a talk on what a heat indicator can and cannot show.
Steam has to reach the liquid, and air inside a sealed bottle expands as it heats. A tight cap can hold the contents below sterilising temperature, or the pressure can crack the bottle or force media out. Back each cap off slightly and tighten it only after the load has cooled.
Fill to the level given in the supplied instructions and check it before every cycle. Too little water can boil away before the holding time ends. The load then heats in dry air instead of steam, and a drum left empty on the heat can be damaged.
Air mixed with steam gives a lower temperature at the same pressure, and pockets of trapped air stay cooler than the steam around them. Let steam push the air out first, as the instructions describe, so the whole load reaches temperature. Packing the drum loosely also helps the steam move between items.
No. The stripes show only that the outside of the pack met steam heat. They say nothing about how long the centre of a dense load stayed hot. Labs that need proof of a cycle use biological indicators, which carry bacterial spores and are incubated after the run.
Leave out sealed containers, flammable solvents and anything that melts or breaks down in moist heat, such as many ordinary plastics. Heat-sensitive solutions such as antibiotic stocks are passed through a sterilising filter instead, and dry glassware can also be treated in a hot-air oven.
Usually the pressure fell too quickly at the end of the run, so the hot liquid boiled up. Switch off and let the vessel cool until the pressure has fallen on its own. Never force the lid or open it under pressure, and leave headspace in every flask and bottle.
To request a quotation, send the product name, code and quantity through our contact page or add the item to your enquiry cart.