Fourier Transform Infrared Spectrometer (FTIR) (Product Code: SCI-304/07) is a single-beam infrared spectrometer used to identify organic and inorganic materials from their infrared absorption spectra in research, quality-control and university laboratories. It measures from 7,800 to 350 cm-1 at a maximum resolution of 0.4 cm-1, using a sealed Michelson interferometer, a Ge/KBr beam splitter and a Peltier-controlled DLaTGS detector. It is supplied by Jain Scientific Equipments Pvt Ltd (ScienceLab), Ambala, India, as one of the laboratory analytical instruments it offers.
| Parameter | Detail |
|---|---|
| Standard measurement range | 7,800 to 350 cm-1 |
| Display wavenumber range | 15,000 to 0 cm-1 (standard) |
| Wavenumber accuracy | Within ± 0.01 cm-1 (theoretical value) |
| Maximum resolution | 0.4 cm-1 |
| Signal-to-noise ratio | 35,000:1 |
| Optical system | Single-beam optics; centre focus, light axis 70 mm high |
| Interferometer | Michelson type (listed at both 30° and 45° incident angle); corner cube mirror with auto-alignment, sealed structure, DSP control |
| Beam splitter | Germanium-coated KBr (Ge/KBr) |
| Light source | High-intensity (high-energy) ceramic source |
| Detector | DLaTGS with Peltier temperature control (standard); MCT (optional) |
| Drive | Mechanical bearing, electromagnetic drive; speed Auto, 1, 2, 3 or 4 mm/sec (DLaTGS 2.0 mm/sec, MCT 4.0 mm/sec) |
| Rapid scan | 10 Hz (optional) |
| Gain switching | Auto, 1, 2, 4, 8, 16, 32, 64, 128 |
| 100%T line flatness | Within 100 ± 1.0 %T (4,000 to 700 cm-1, continuous repetitive measurement) |
| Sample chamber | 200 mm (W) x 260 mm (D) x 185 mm |
| Nitrogen purge | Interferometer, sample compartment and detector |
| Microscope functions listed | High-sensitivity MCT detector; auto aperture, auto centering, auto X-Y stage and auto focus; memory for up to 10 sample and 2 background positions; control from the computer screen or the microscope's own keyboard |
| Communication | USB 2.0 |
| Product code | SCI-304/07 |
| Supplied by | Jain Scientific Equipments Pvt Ltd (ScienceLab), Ambala, India |
An FTIR spectrometer shows which chemical bonds and functional groups a sample contains, so it is used to identify unknown substances, confirm the identity of pharmaceutical and polymer raw materials, check product purity and follow reactions. The interferometer collects all wavelengths at once and a Fourier transform converts the result into a spectrum that can be compared with reference spectra. In universities it is a core instrument for organic chemistry, materials science and research projects.
Install the spectrometer in a dry, temperature-stable room, because KBr optics are sensitive to humidity. Use the nitrogen purge to reduce water-vapour and carbon-dioxide bands, and record a fresh background before each sample series. Keep the sample compartment clean and closed when the instrument is idle.
Jain Scientific Equipments Private Limited, trading as ScienceLab, supplies this FTIR spectrometer from Ambala, Haryana, to universities, research institutes, testing laboratories, dealers and importers. We confirm the configuration with you, prepare the quotation and organise packing and shipment. Company details are on our profile page.
Instruments and accessories often used alongside infrared spectroscopy:
Explore the wider Laboratory Equipment category for more instruments.
What wavenumber range and resolution does the FTIR spectrometer cover?
The Fourier Transform Infrared Spectrometer (SCI-304/07) has a standard measurement range of 7,800 to 350 cm-1, a display range of 15,000 to 0 cm-1, maximum resolution of 0.4 cm-1 and wavenumber accuracy within ± 0.01 cm-1 (theoretical value).
Which detector is fitted to the FTIR spectrometer?
The FTIR spectrometer is fitted as standard with a DLaTGS detector under Peltier temperature control, run at 2.0 mm/sec. An MCT detector, run at 4.0 mm/sec, is listed as optional for higher-sensitivity work; confirm your choice at quotation stage.
Is an infrared microscope included with the FTIR spectrometer?
The FTIR listing describes microscope functions such as auto aperture, auto centering, auto X-Y stage, auto focus and memory for 10 sample and 2 background positions. Please confirm with us whether the microscope unit is part of the configuration you are quoted.
How large is the FTIR sample compartment?
The FTIR spectrometer's sample chamber measures 200 mm wide by 260 mm deep by 185 mm, with a centre-focus optical path and the light axis 70 mm high. The interferometer, sample compartment and detector can all be purged with nitrogen.
How does the FTIR spectrometer connect to a computer?
The FTIR spectrometer communicates over USB 2.0. The listed microscope operations, including stage movement, aperture setting, focusing and switching between transmission and reflection modes, can be controlled from the computer screen or from the microscope's own keyboard.
Who supplies the FTIR spectrometer?
The Fourier Transform Infrared Spectrometer (FTIR) is supplied by Jain Scientific Equipments Pvt Ltd (ScienceLab), Ambala, India, to universities, research and testing laboratories in India and international markets. Ask us which make and model is currently offered.
Can a calibration certificate be provided with the FTIR spectrometer?
Whether a calibration certificate can accompany the FTIR spectrometer should be discussed with us before ordering, because it depends on the model quoted. Mention it in your enquiry through our contact page along with your detector choice.
For a quotation on the FTIR spectrometer (SCI-304/07), write to us via the contact page or place it in the enquiry cart. Mention your samples, detector preference and whether the microscope functions are needed.
Last updated: 24 September 2026