Small reagent quantities in student chemistry and biology practicals are weighed on the Compact Balance 300g, a portable centigram balance. It reads up to 300 g with 0.01 g readability, 0.01 g repeatability and 0.02 g linearity on a 100 mm removable pan. Three colour-coded buttons sit on a sealed keypad, and power comes from the included adapter or from 6 AA batteries that the user supplies.
| Parameter | Detail |
|---|---|
| Capacity and readability | 300 g; 0.01 g |
| Repeatability and linearity | 0.01 g; 0.02 g |
| Stabilisation time | 2.0 (as listed) |
| Pan | 100 mm, removable platform |
| Draught shield | Fitted |
| Units and calibration | g, oz; external calibration, initiated from the keypad |
| Display, keypad and software | High-contrast, 15 mm digit height; three colour-coded buttons, sealed against dirt and spills; auto power-off; zero tracking; overload and low-battery indicators |
| Power | 12V AC 50/60Hz 150mA adapter (included) / 6 x AA batteries (not included); battery life 3 hours |
| Construction and size | ABS plastic, with overload protection; net weight 0.5 kg; 147 x 208 x 150 mm (W x D x H); operating temperature 0° to 40°C |
| Product code | SLE/LB/23 |
| Minimum order | 10 units |
| Dispatch | Usually within 7 to 10 working days |
| Supplied by | ScienceLab (Jain Scientific Equipments Pvt Ltd), Ambala, India |
Rate of reaction by mass loss. Stand a conical flask of dilute hydrochloric acid on the pan, add calcium carbonate granules and plug the neck loosely with cotton wool. Record the mass at the start and then at fixed intervals timed with a digital stopwatch. Repeat with powder in place of granules to see how surface area changes the rate.
Solution of known concentration. Tare a sheet of weighing paper, then add sodium chloride with a polypropylene spatula until the display shows the mass needed to two decimal places. Tip the salt through a funnel into a volumetric flask, rinse the paper and funnel into it, and make up to the mark with distilled water.
Water loss from leaves. Pick four similar leaves and coat them with petroleum jelly on the upper side, the lower side, both sides or neither. Weigh each leaf, hang them on a line for a few days and weigh them again. The leaves with a bare lower surface lose the most mass, which shows where most stomata lie.
Treat the final digit as uncertain by about one step. With 0.01 g repeatability, repeat weighings of one sample may differ in the second decimal place. Record masses to two decimal places, and remember that a sample of only a few hundredths of a gram carries a large percentage uncertainty.
Repeatability of 0.01 g is how closely repeat readings of the same load agree. Linearity of 0.02 g is how far a reading may stray from the true value at different points between zero and 300 g. In the worst case both add together, so allow a few hundredths of a gram on any one result.
The loose plug lets carbon dioxide escape but catches the fine spray of acid that the fizzing throws up. The mass lost is then due to the gas alone, which is what the experiment measures. It also keeps acid droplets off the pan and the sealed keypad.
That reaction gives off hydrogen, which is so light that the total mass lost is tiny, often only a few hundredths of a gram. The change is close to the reading uncertainty, so the results scatter. Collect the hydrogen in a gas syringe instead and measure its volume.
An electronic balance senses the force of the load, so local gravity, temperature and the slope of the bench all shift its reading slightly. External calibration with a known mass, started from the keypad at the place of use, corrects for them. A calibration mass is not stated as included.
Rate from mass loss is given in grams per second or grams per minute. A mean rate is the total mass lost divided by the time taken. For the rate at one moment, draw a tangent to the curve of mass against time and work out its gradient.
To request a quotation, send the product name, code and quantity through our contact page or add the item to your enquiry cart.