The Balance, Analytical is a four-decimal electronic balance that weighs up to 220 g with a resolution of 0.0001 g. It uses an electromagnetic force restoration weighing system with internal calibration, and a large glass draught shield with 3 sliding doors encloses the 80 mm pan. A backlit LCD, RS232 interface and a menu of application functions complete it.
| Parameter | Detail |
|---|---|
| Capacity | 220 g |
| Resolution | 0.0001 g |
| Weighing system and calibration | Electromagnetic force restoration; internal calibration |
| Draught shield | Large glass shield with 3 sliding doors |
| Pan size | Ø 80 mm |
| Display and interface | LCD with backlight; RS232 |
| Housing | Double case structure; ergonomic keypad; plastic protection cover |
| Units of measure | Grams, milligrams, carats, ounces, pounds, pennyweights, ounce troy, grain, pieces |
| Application functions | Selectable filter level; selectable units; piece counting; density; peak hold; under hook; checkweighing; percentage weighing; add weighing; animal weighing |
| Product code | SLE/LB/18 |
| Minimum order | 10 units |
| Dispatch | Usually within 7 to 10 working days |
| Supplied by | ScienceLab (Jain Scientific Equipments Pvt Ltd), Ambala, India |
Primary standard for acid titration. Dry sodium carbonate, cool it in a vacuum desiccator and keep it in a stoppered bottle. Place the bottle on the pan, tare, remove a portion with a Chattaway spatula into a titration flask and read the negative value, which is the mass taken. Titrate it against dilute hydrochloric acid to find the acid's exact concentration.
Uniformity of tablet mass. Weigh a batch of tablets together and work out the mean mass, then weigh each one alone and record it. Students work out how far each tablet differs from the mean, and send the readings through the RS232 interface to a printer or computer so the record matches the balance display.
Routine performance check. Place a reference mass from the analytical balance mass set at the centre of the pan and then towards the front, back and sides, and compare the readings. Repeat one mass ten times for repeatability. Stand the balance on a balance anti-vibration table if readings wander with movement in the room.
In daily use, yes. The weighing cell and electronics give the steadiest readings once they have settled at a stable temperature, so leave the balance connected to the mains. After a cold start, allow the warm-up time the operating instructions give before calibrating or weighing.
Taring sets the display to zero with the container on the pan, so only the sample is shown. Internal calibration uses the built-in reference to correct the balance across its range. Tare before every weighing, and run the internal calibration after installation, after a move and after a large temperature change.
Wait until the last digit stops changing with all doors closed. The filter level setting decides how strongly the balance smooths disturbances: a higher level steadies the display in a busy room but responds more slowly. Record the value straight away and do not round it.
Plastic containers and dry powders pick up charge, especially in dry air, and the reading then drifts or jumps. Use glass or metal containers where possible, touch the container to an earthed metal surface before weighing and keep the room from becoming very dry.
At 0.0001 g resolution, air moving through an open door presses on the pan enough to change the last digit. Open only the door needed to load the sample, close it again and wait for the display to settle. Close all 3 sliding doors after use as well.
No. Resolution is the smallest step the display shows. Real accuracy also depends on calibration, repeatability, the room and how the sample is handled. A very small sample carries a large relative error, so weigh enough material that the last digit matters little.