The Table Top Balance weighs kilogram-scale loads on a 210 x 260 mm stainless steel platter. It comes in 3 kg, 10 kg and 30 kg capacities, with readability from 0.1 g to 2 g depending on the model. A built-in rechargeable battery lets it work away from a socket, and readings show on a red LED display.
| Parameter | Detail |
|---|---|
| Capacity options | 3 kg, 10 kg, 30 kg |
| Readability options | 0.1 g, 0.5 g, 1 g or 2 g, depending on model |
| Resolution | Up to 30,000 counts |
| Pan size | 210 x 260 mm |
| Platter | Rust-proof stainless steel |
| Display | High-brightness red LED |
| Power | Built-in rechargeable battery; battery indicator; auto power off |
| Interface | RS-232C (optional) |
| Protection | Overload protection design |
| Product code | SLE/LB/11 |
| Minimum order | 10 units |
| Dispatch | Usually within 7 to 10 working days |
| Supplied by | ScienceLab (Jain Scientific Equipments Pvt Ltd), Ambala, India |
Water content of soil. Weigh an empty porcelain dish, add fresh soil taken with a soil auger and weigh again. Dry the dish in a laboratory oven, let it cool in a desiccator and reweigh. The mass lost is the water, which is then expressed as a percentage of the dry soil mass.
Sieve analysis of sand. Weigh a dry sand sample, shake it through a stack of test sieves and weigh the portion held on each sieve. The retained masses should add up close to the starting mass; a large shortfall means sand was spilled or left trapped in the mesh.
Off-centre check. Put a known mass from a weight box at the centre of the platter, note the reading, then move it to each corner in turn. If the readings differ, centre bulky loads such as a bucket or a large polypropylene beaker on the platter every time.
Look at the battery indicator first and charge the built-in battery fully if it is low. A display that goes blank after a spell without use is usually the auto power off, not a fault. If it still will not start, inspect the charger lead for breaks or loose plugs.
The stainless steel platter does not rust, but water and grit can run off its edges into the body and reach the load cell and electronics. Weigh wet material in a tray or beaker, tare the container first, and wipe the platter with a damp cloth, then dry it, after the session.
A warm tray sets up rising air currents that make the reading low and unsteady. Dry soil left to cool on an open bench also takes moisture back from the air, which raises its mass. Cooling it in a closed desiccator avoids both errors.
Return the tray to the oven, cool it and weigh it again. When two weighings in a row agree, the soil has reached constant mass. Stopping too early leaves water in the soil, so the calculated water content comes out lower than it really is.
Damp sand clumps together, so fine grains cling to coarse ones and stay on the upper sieves. Its water also adds mass that disappears as the sand dries during shaking. Dry the sample, let it cool, then weigh it before it goes through the sieves.
To request a quotation, send the product name, code and quantity through our contact page or add the item to your enquiry cart.